Mechanical ventilation is a vital mainstay in the management of patients with respiratory failure in intensive care units and operating theatres. The choice of ventilation mode and the setting of its parameters depend on an understanding of the mechanical properties of the lung and the physiological mechanisms that ensure effective gas exchange while minimising pulmonary risks.
Volume-Controlled and Pressure-Controlled Ventilation
Volume control delivers a constant inspiratory flow to reach a pre-set tidal volume, which ensures constant minute ventilation, but the inflation pressure within the airways rises if resistance increases or lung compliance falls. In contrast, Pressure control sets a target pressure at which flow starts high and then gradually decelerates, which limits peak alveolar pressure with the aim of reducing the risk of ventilator-induced lung injury (VILI), but its main drawback is that the tidal volume is not guaranteed and varies in proportion to changes in compliance.
Supportive and Synchronised Modes: SIMV and Pressure Support
SIMV is characterised by synchronising mandatory breaths with the patient's respiratory effort, which prevents the phenomenon of breath stacking, while allowing the patient to take free spontaneous breaths between these mandatory breaths. Pressure support, on the other hand, depends entirely on patient triggering, with each breath initiated by the patient supported by a pre-set pressure (usually ranging between 5–20 cmH2O), and the breath ends as soon as flow falls below a certain level, which reduces the work of breathing as support is increased.
Continuous Positive Pressure and Complications of Mechanical Ventilation
CPAP or the application of PEEP provides continuous positive pressure throughout the respiratory cycle (usually between 5–20 cmH2O), which keeps the airways open, increases functional residual capacity (FRC) and reduces atelectasis without adding tidal volume. The device senses the patient's effort (triggering) by a change in flow (usually 2 L/min) or pressure (2 cmH2O). Among the most important complications is auto-PEEP, resulting from insufficient expiratory time, and it is treated by prolonging expiration or by temporarily disconnecting the patient in an emergency. VILI also includes volutrauma, atelectrauma and biotrauma, in addition to barotrauma caused by excessive pressures.
- Volume control ensures constant volume and minute ventilation but may raise pressures when compliance falls.
- Pressure control limits peak alveolar pressure to reduce VILI, but the tidal volume fluctuates with changes in compliance.
- Pressure support reduces the work of breathing, and its inspiratory phase ends when flow falls to a specified level.
- PEEP and CPAP increase functional residual capacity (FRC) and prevent atelectasis without delivering additional tidal volume.
- Auto-PEEP results from a short expiratory time, and is treated by prolonging the expiratory period or temporarily disconnecting the patient in an emergency.