Anaesthetic vaporisers are among the most important components of the anaesthetic workstation for converting liquid drugs into therapeutic vapours with extreme precision. The safe performance of these vaporisers depends on a precise understanding of bypass mechanisms, physical compensation, and the proper handling of environmental variables and common errors.
The variable bypass principle and thermal compensation
The variable bypass principle relies on dividing the fresh gas flow into two streams: one passes through the vaporising chamber and becomes fully saturated, and the other passes through the bypass stream, to meet again at the outlet. The splitting ratio is defined as the ratio of bypass flow to vaporising chamber flow. To counter the cooling caused by the latent heat of vaporisation, the vaporiser contains a metal heat sink to absorb heat and a bimetallic strip or bellows that directs more gas to the vaporising chamber as the temperature falls.
The pumping effect, the properties of Desflurane and the Tec 6 vaporiser
Back pressure resulting from intermittent positive pressure ventilation (IPPV) or operating the oxygen flush button (O2 flush) may increase vaporiser output, and this phenomenon is known as the pumping effect, which reaches its maximum at low flows, low dial settings and high inspiratory pressures. On the other hand, Desflurane has a saturated vapour pressure reaching 669 mmHg and boils at 22.8°C, so it requires a special vaporiser (Tec 6) that heats it to 39°C under pressure to ensure its controlled output.
The effect of altitude and operating hazards
At high altitudes, a variable bypass vaporiser delivers a higher volume percentage (vol%) with unchanged partial pressure, which leaves the clinical depth unchanged. In contrast, the Tec 6 vaporiser delivers a constant volume percentage, so the partial pressure falls with altitude, which requires the dial setting to be raised. Among the most prominent operating hazards is filling the vaporiser with a liquid whose saturated vapour pressure is higher than its calibration, leading to an excessive dose that may be fatal in the case of Desflurane. Also, tipping of the vaporiser allows liquid to enter the bypass pathway; its treatment is to flush it with a high flow at a low dial setting.
- The splitting ratio is the ratio of bypass flow to vaporising chamber flow.
- Thermal cooling is countered by a metal heat sink and a bimetallic strip that directs more gas to the chamber when the temperature falls.
- The pumping effect peaks with low flows, low settings and high inspiratory pressures.
- The Tec 6 vaporiser heats liquid Desflurane to 39°C under appropriate pressure because it boils at 22.8°C.
- Tipping of the vaporiser leads to liquid entering the bypass pathway and is treated with high flow at a low dial setting.
Sources
- Anesthesia Equipment: Principles and Applications, pp. 70–71
- Anesthesia Equipment: Principles and Applications, p. 71
- Understanding Anesthesia Equipment, pp. 125–126
- Anesthesia Equipment: Principles and Applications, pp. 75–76
- Spotlights on Anesthesia, p. 393
- Anesthesia Equipment: Principles and Applications, p. 80
- Spotlights on Anesthesia, p. 394
- Anesthesia Equipment: Principles and Applications, pp. 72–73
- Understanding Anesthesia Equipment, p. 185