Oxygen monitoring is a fundamental pillar of ensuring patient safety during anaesthesia and surgical operations. These techniques rely on precise physical and chemical principles that ensure the efficiency of oximetry devices and oxygen analysers.
Physical principles of the pulse oximeter
Measurement of light absorbance relies on two main laws; Beer's law states that transmitted light decreases exponentially with the concentration of the absorbing substance, whereas Lambert's law states that it decreases exponentially with the length of the light path. The pulse oximeter applies these two principles using two wavelengths: red light at 660 nm and infrared at 940 nm. Deoxyhaemoglobin absorbs light more at 660 nm, whereas oxyhaemoglobin absorbs light more at 940 nm.
Influencing factors and sources of error in readings
The accuracy of the pulse oximeter is affected by several clinical factors; in carbon monoxide poisoning, carboxyhaemoglobin is read to a large extent as if it were oxyhaemoglobin, giving a falsely high SpO2 reading. In the case of methaemoglobin, it absorbs equally at 660 and 940 nm, which causes readings to trend towards 85%. In addition, injection of methylene blue dye or the presence of dark nail polish causes falsely low readings, while poor peripheral perfusion and movement lead to failure to pick up the signal and the appearance of measurement artefact.
Oxygen analysers: paramagnetic versus fuel cell
Two types of device are used to measure oxygen concentration in anaesthesia circuits; the paramagnetic analyser is distinguished by the unpaired electrons in the oxygen molecule that make it attracted to a magnetic field, providing a very rapid response that allows breath-by-breath oxygen measurement. In contrast, the fuel cell (galvanic) analyser works as a chemical battery that needs no external power source, as it contains a gold cathode and a lead anode. However, it suffers from a relatively slow response of about 20 seconds, and a limited operating life of about one year because of consumption of the lead within it.
- The pulse oximeter measures light absorption at 660 nm for deoxyhaemoglobin and at 940 nm for oxyhaemoglobin, based on Beer's and Lambert's laws.
- Carboxyhaemoglobin gives falsely high SpO2 readings, whereas methaemoglobin drives the reading towards 85%.
- Injected methylene blue and dark nail polish cause falsely low readings, and poor perfusion and movement cause signal disturbance.
- The paramagnetic analyser has a very rapid response for breath-by-breath oxygen analysis, relying on unpaired electrons.
- The fuel cell analyser works as a battery without external electricity but responds slowly and has a life limited to about a year because of consumption of the lead anode.
Sources
- Essentials of Anaesthetic Equipment, 4th ed., pp. 151–154
- Essentials of Anaesthetic Equipment, 4th ed., p. 153
- Essentials of Anaesthetic Equipment, 4th ed., pp. 158–160
- Essentials of Anaesthetic Equipment, 4th ed., p. 159