Modern anaesthetic practice depends on a continuous and safe supply of the essential medical gases within operating theatres. This system requires a precise understanding of cylinders, pipeline networks, and central storage sources to ensure patient safety and prevent accidents.
Colour Coding and Safety Systems in Cylinders
Gas cylinders are distinguished according to two colour-coding systems; the old British system (old UK coding) gives oxygen a black body with a white shoulder, and nitrous oxide a blue body and shoulder, while a medical air cylinder is distinguished by a grey body and a shoulder divided into quarters of white and black, and Entonox gas has a blue body and a shoulder divided into quarters of blue and white. The new system, on the other hand, adopts white bodies for all cylinders with distinctive colours on the shoulders. To prevent error at connection, the Pin Index Safety System is applied to small cylinders through unique pin positions: oxygen occupies position 2,5, nitrous oxide occupies 3,5, and air occupies 1,5, while position 7 is allocated to Entonox gas.
Physical Properties and Monitoring of Contents
The physical properties of the gases inside the cylinders differ fundamentally, which affects monitoring of the quantities remaining. A full oxygen cylinder is at a pressure of 13700 kPa and contains pure gas, which makes the pressure gauge reading directly proportional to the actual contents. In contrast, a full nitrous oxide cylinder has a pressure of 4400 kPa and contains a mixture of liquid and vapour; as a result, the pressure remains completely constant as long as there is liquid vaporising inside the cylinder. Accordingly, the pressure gauge does not reflect the quantity of nitrous oxide accurately, and one must rely solely on the weight of the cylinder to determine its contents.
The Pipeline Network and VIE Units
Medical facilities supply gases through the pipeline network (piped medical gases) at an operating pressure of about 400 kPa (equivalent to 4 bar), while the air intended to power surgical tools (surgical tools) is pumped at a higher pressure of up to 700 kPa. In large hospitals, liquid oxygen is stored in a vacuum-insulated evaporator (Vacuum insulated evaporator - VIE) at extremely cold temperatures ranging between −150 and −170°C inside a vacuum wall. This system contains superheater coils (superheater coils) to vaporise the liquid during periods of high consumption, and a safety valve (safety valve) to release excess pressure, and its contents are measured by weight or by measuring the pressure difference.
- Pin Index positions prevent mix-ups: oxygen occupies 2,5, nitrous oxide occupies 3,5, air occupies 1,5, and Entonox occupies 7.
- The pressure of a full oxygen cylinder is 13700 kPa and is directly proportional to its contents, whereas a full nitrous oxide cylinder has a pressure of 4400 kPa and its contents are measured by weight.
- The pipeline network supplies medical gases at a pressure of 400 kPa (4 bar), while surgical tool air is supplied at a pressure of 700 kPa.
- The VIE system stores liquid oxygen between −150 and −170°C, and uses superheater coils for vaporisation and a safety valve to release excess pressure.
Sources
- Essentials of Equipment in Anaesthesia, 5th ed., pp. 4–5
- Essentials of Equipment in Anaesthesia, 5th ed., p. 7
- Essentials of Equipment in Anaesthesia, 5th ed., p. 10