Theatre Ventilators: Principles, Modes and Safety Measures

Topic notes · Equipment & Monitoring · The anaesthetic machine and gases

Ventilators in operating theatres are a fundamental pillar for ensuring stable gas exchange and oxygenation in patients undergoing general anaesthesia. Safe use of these devices requires a precise understanding of the characteristics of ventilation modes, the physiological effects of pressures, and correct setting of the alarm system to protect the patient from any pulmonary or circulatory injury.

Ventilation modes and applying a lung-protective strategy

Volume-controlled ventilation depends on delivering a set and constant tidal volume, while airway pressure varies according to changes in lung compliance and airway resistance. In contrast, pressure-controlled ventilation sets the target inspiratory pressure while the tidal volume delivered to the patient varies according to the properties of their respiratory system. To apply lung-protective ventilation, it is recommended to set the tidal volume between 6 and 8 mL/kg based exclusively on ideal or predicted body weight, with application of positive end-expiratory pressure (PEEP) and recruitment manoeuvres. Predicted body weights are calculated by standard equations: for males [50 + 0.91 × (height in centimetres − 152.4)], and for females [45.5 + 0.91 × (height in centimetres − 152.4)].

PEEP physiology and the inspiratory to expiratory ratio

Applying PEEP achieves important clinical benefits; it raises the mean airway pressure, restores functional residual capacity (FRC), and reduces the occurrence of atelectasis, which contributes to improved oxygenation. However, excessive PEEP should be avoided, because it may cause alveolar overdistension and barotrauma, in addition to reduced venous return and the consequent hypotension. On the other hand, the usual inspiratory to expiratory ratio (I:E ratio) in adults is 1:2 or a longer expiratory time; when faced with the phenomena of auto-PEEP or breath stacking, they are corrected clinically by prolonging expiratory time.

The pressure alarm system and patient safety

Alarms form the critical safety line for detecting mechanical and physiological disturbances during surgery. The low-pressure alarm warns of disconnection of the anaesthetic circuit or the presence of a gas leak when peak pressure falls below the threshold, and safety requires setting this threshold very close to the peak pressure. The high-pressure alarm, on the other hand, serves to alert the medical staff to the presence of airway obstruction, bronchospasm, accumulated secretions, patient coughing, or a sudden fall in lung compliance.

Key points
  • In VCV the tidal volume is constant and the pressure variable, whereas PCV sets the inspiratory pressure and the tidal volume is variable.
  • Lung-protective ventilation depends on a tidal volume of 6–8 mL/kg of predicted body weight (PBW) with application of PEEP and recruitment manoeuvres.
  • PEEP improves oxygenation and FRC, but excess leads to barotrauma, reduced venous return and hypotension.
  • The normal I:E ratio in adults is 1:2 or a longer expiratory time, and correcting auto-PEEP and breath stacking requires prolonging expiratory time.
  • The low-pressure alarm detects disconnection and leaks, whereas the high-pressure alarm warns of bronchospasm, obstruction and reduced compliance.

Sources

  • Clinical Anesthesia, 9th ed., p. 744
  • Clinical Anesthesia, 8th ed., p. 709
  • Clinical Anesthesia, 9th ed., p. 356
  • Pilbeam's Mechanical Ventilation, 7th ed., p. 71
  • Essentials of Anaesthetic Equipment, 4th ed., p. 164
  • Essentials of Anaesthetic Equipment, 4th ed., p. 165