ASDr. Ahmed ShihabConsultant Anesthesiologist – Baghdad عربي

Electrical Safety, Diathermy and Operating Theatre Fires

Topic notes · Equipment & Monitoring · Other equipment and safety

Electrical safety and fire prevention in the operating theatre are fundamental pillars that ensure the safety of the patient and the surgical team. Daily handling of monitoring equipment and Diathermy tools requires a precise understanding of the principles of electrical isolation and the causes of ignition in order to avoid catastrophic incidents.

Risks of Electric Shock and Classification of Equipment

Electric shocks fall into two main types; Macroshock, which occurs when current passes across the body surface, where the patient feels about 1 mA, and the let-go threshold reaches 10–20 mA, whereas a current of 100–300 mA causes ventricular fibrillation (VF) at a frequency of 60 Hz for one second. Microshock, on the other hand, occurs when current is applied directly to the myocardium, where an intensity of only 50–100 µA is enough to cause ventricular fibrillation, so the recommended maximum leakage current has been set at 10 µA. In terms of equipment classification, Class I devices rely on connecting the outer casing to a protective earth, whereas Class II devices rely on double insulation without earthing. As for the patient-applied parts, Type CF parts are the only ones safe for direct cardiac connection because they have the lowest leakage current, unlike Type B and Type BF, which are not cardiac-safe.

The Ungrounded Power System and Insulation Monitoring

Operating theatres are supplied with a special power system based on an isolation transformer that provides ungrounded power, which means that contact of a single live line with earth does not complete the electrical circuit and does not cause a shock. To monitor the efficiency of this system, a Line Isolation Monitor is used; its function is limited to sounding a warning alarm when a fault in the insulation is detected and leakage exceeds the permitted limit, without cutting off the electrical current, in order to ensure the continued operation of vital and life-supporting equipment.

Diathermy and the Airway Fire Protocol

The electrosurgical unit (Diathermy) uses a radiofrequency current that heats tissue and cuts or coagulates it without stimulating the heart. In the Monopolar system, a high current density is concentrated at the active tip and then returns through a large plate, and any fault or poor contact of this plate causes severe burns to the patient. In the Bipolar system, by contrast, the current passes only between the two forceps tips with no need for a return plate. The use of diathermy is associated with the risk of fires through completion of the fire triangle composed of oxidisers (O2 and N2O), fuel (the tracheal tube and drapes), and an ignition source (the Diathermy or laser). In the event of an airway fire, surgery and gas flow must be stopped immediately, the tube and burning materials removed, the area flooded with saline, and then ventilation with air given, followed by re-intubation and an endoscopic examination with the bronchoscope.

Key points
  • The let-go threshold in Macroshock is 10–20 mA, whereas 50–100 µA is enough in Microshock to cause ventricular fibrillation, and the maximum permitted leakage is 10 µA.
  • Class I devices are connected to protective earth, whereas Class II is characterised by double insulation, and Type CF parts alone are considered safe for direct connection to the heart.
  • The isolation transformer provides ungrounded power, while the LIM sounds an alarm when leakage occurs without cutting off the electricity.
  • Monopolar diathermy needs a return plate to prevent burns, whereas the Bipolar current is confined between the two forceps tips.
  • Immediate management of an airway fire includes stopping surgery and gases, removing the tube, extinguishing with saline, and ventilating with air before re-intubation and using the bronchoscope.

Sources

  • Clinical Anesthesia, p. 332
  • Clinical Anesthesia, pp. 346–348
  • Essentials of Equipment in Anaesthesia, 5th ed., pp. 247–248
  • Essentials of Equipment in Anaesthesia, 5th ed., p. 248
  • Essentials of Equipment in Anaesthesia, 5th ed., pp. 251–252
  • Emergency Management in Anaesthesia, p. 103

Test yourself: 7 questions on this topic and many more →