Breathing systems and carbon dioxide absorbers are among the fundamental pillars of airway and ventilation management during general anaesthesia. A precise understanding of the characteristics of these systems and their chemical reactions ensures patient safety and avoids the accumulation of harmful gases or toxins within the circuit.
Mapleson systems and the differences in their breathing efficiency
Mapleson systems differ in efficiency according to the ventilation pattern and the fresh gas flow (FGF). The Mapleson A system, such as the Magill and Lack, is the most efficient during spontaneous ventilation, requiring a flow equal to the minute volume (70–100 ml/kg/min), but it becomes the least efficient during controlled ventilation, as it requires about three times the minute ventilation. The Mapleson D system, known as the Bain, by contrast is highly efficient in controlled ventilation at a flow of 70–100 ml/kg/min, and requires 150–200 ml/kg/min for spontaneous breathing. The general order of efficiency in spontaneous breathing is: A > D, F, E > C, B, and in controlled ventilation: D, F, E > B, C > A. The Mapleson C system is used in resuscitation and manual ventilation (resuscitation/bagging), whereas the Mapleson F system, known as the Jackson Rees, is reserved for children weighing less than 25 kg.
Structure of the circle system and properties of soda lime
The circle system consists of main components including the soda lime canister, the fresh gas inlet, inspiratory and expiratory limbs fitted with unidirectional valves, a Y-piece, an APL valve, and a gas collecting reservoir bag. Sticking or inefficiency of the unidirectional valves leads to rebreathing of CO2 and hypercapnia. Soda lime contains about 94% calcium hydroxide, 5% sodium hydroxide, and 14–19% water, in addition to silica to harden the granules and prevent their disintegration. The indicator Ethyl violet is added, which changes colour from white to violet as a sign that the absorbent's capacity has been consumed.
Chemical reactions and toxic hazards of absorbents
Reactions between anaesthetic gases and absorbents may lead to undesirable by-products. Compound A is formed when sevoflurane is broken down by the strong bases in the absorbent, and the chance of its formation increases when working with a low fresh gas flow, high concentrations of sevoflurane, and a rise in temperature. On the other hand, the dangerous gas Carbon monoxide is formed when desflurane (and to a lesser degree isoflurane) passes through a completely desiccated absorbent, and this toxic reaction is avoided by using alkali-free absorbents.
- The Mapleson A system is the most efficient in spontaneous breathing, whereas Mapleson D excels in controlled ventilation.
- The Mapleson C system is used in resuscitation, and the Mapleson F system is reserved for children below a weight of 25 kg.
- Damage to or inefficiency of the unidirectional valves in a circle system leads to CO2 rebreathing and hypercapnia.
- The indicator Ethyl violet changes from white to violet as a sign that the soda lime has been consumed.
- Compound A results from the breakdown of sevoflurane at low flow, whereas Carbon monoxide results from the reaction of desflurane with dry absorbent.
Sources
- Smith and Aitkenhead's Textbook of Anaesthesia, p. 309
- Essentials of Equipment in Anaesthesia, p. 62
- Morgan & Mikhail's Clinical Anesthesiology, p. 38
- Basics of Anesthesia, p. 248