Monitoring of the circulation and the electrocardiogram are two essential pillars for ensuring patient safety during anaesthesia in the operating theatre. This article aims to clarify the accepted physical and clinical principles for setting up blood pressure measuring devices and the ECG with high accuracy.
Non-invasive blood pressure measurement (NIBP)
Non-invasive blood pressure measurement with Oscillometric NIBP devices relies on measuring arterial oscillations; the mean arterial pressure (MAP) corresponds to the maximum oscillation amplitude (maximum oscillation). The rapid rise in oscillations indicates the systolic pressure, while the rapid fall represents the diastolic pressure. To achieve accurate measurements, the cuff bladder width must be 40% of the limb circumference; a small cuff gives falsely high readings (overestimates BP), whereas a large cuff gives falsely low readings (underestimates BP).
Invasive arterial pressure monitoring
The invasive arterial pressure measurement system consists of a cannula, a saline fluid column, a flush device, a transducer, an amplifier, and a display. Accuracy requires zeroing, by opening the transducer to air to cancel the effect of atmospheric pressure, in addition to levelling by placing the transducer at the phlebostatic axis in the fourth intercostal space at the mid-axillary line. Placing the transducer too low leads to falsely high readings (over-reads), whereas placing it too high leads to falsely low readings (under-reads).
Fluid dynamics and damping phenomena
The natural frequency of the monitoring system must be much higher than the pulse rate to avoid resonance. The natural frequency rises with increasing catheter diameter, and falls with the use of longer tubing or more compliant systems. In clinical situations, overdamping, caused by air bubbles, clots or compliant tubing, results in a blunted waveform, an absent dicrotic notch, and a falsely low systolic pressure. In contrast, underdamping leads to systolic overshoot and a falsely high systolic pressure reading.
Electrocardiogram (ECG) monitoring
Lead V5 is the most sensitive single lead for detecting myocardial ischaemia, and detection accuracy increases when it is combined with lead II (II + V5). For arrhythmias, lead II is the optimal and most suitable choice, because its electrical axis parallels the P wave axis, which facilitates accurate diagnosis of atrial activity.
- MAP corresponds to the maximum oscillation in an Oscillometric NIBP device, and the cuff width must cover 40% of the limb circumference to avoid errors.
- Invasive pressure measurement requires placing the transducer at the phlebostatic axis (4th intercostal space, mid-axillary line) to prevent erroneous readings.
- Overdamping leads to a blunted waveform, an absent dicrotic notch and a lowered systolic pressure, whereas underdamping causes systolic overestimation.
- The natural frequency falls with increasing tubing length and compliance, exposing the system to the phenomenon of resonance.
- Lead V5 is the most sensitive single lead for detecting ischaemia, whereas lead II is the best for arrhythmias because it parallels the P wave axis.
Sources
- Essentials of Equipment in Anaesthesia, 6th ed., Ch. 10
- Essentials of Equipment in Anaesthesia, 6th ed., Ch. 11
- Miller's Anesthesia, 9th ed., p. 1161
- Miller's Anesthesia, 9th ed., p. 1162
- Clinical Anesthesia, 8th ed., p. 397