- The pulse oximeter (SpO₂) measures oxygenation; the capnograph (EtCO₂) measures ventilation. With supplemental oxygen the first lags and the second leads.
- The normal capnogram has four phases and two angles: alpha about 100–110 degrees, beta about 90 degrees.
- The normal EtCO₂ range is 35–45 mmHg, about 2–5 mmHg below arterial PaCO₂ in a healthy person.
- Few, long, tall waves = hypoventilation. Many, short, low waves = hyperventilation. Same shape; what changed is height and rate.
This page is the reference version of the «Capnography Decoded» (الموجة تحچي) series published on Instagram. New episodes are added here as they are published.
Episode 1 — Oxygen 99%… and your patient is not breathing
The principle: two devices measure two different things
Pulse oximetry (SpO₂) measures oxygenation – haemoglobin saturation with oxygen in peripheral blood. Capnography (EtCO₂) measures ventilation – how effectively carbon dioxide is eliminated with each breath.
Why does the oximeter lag?
A patient receiving supplemental oxygen has an oxygen store in the functional residual capacity (FRC), so saturation stays normal for a while even after ventilation stops – delayed desaturation. Capnography reads breath by breath, so it catches hypoventilation and apnoea before saturation falls.
Where does this matter clinically?
- Procedural sedation: the ASA standards require monitoring of exhaled carbon dioxide during moderate and deep sedation.
- PACU: a patient on oxygen under the effect of anesthetic and analgesics.
- Opioid-induced respiratory depression: opioids reduce ventilation while SpO₂ stays good as long as the patient is on supplemental oxygen.
- Airway obstruction: the waveform changes or disappears.
Episode 2 — Every breath has four phases
The normal capnogram is not a line but the story of a whole breath:
| Phase | Name | What happens |
|---|---|---|
| I | Baseline | Dead-space gas that takes no part in gas exchange, so it contains no carbon dioxide. |
| II | Expiratory upstroke | Alveolar gas begins to mix with dead-space gas. |
| III | Alveolar plateau | Pure alveolar gas. Its last point is the number on the screen: EtCO₂. |
| IV | Inspiratory downstroke | Inspiration: fresh gas enters and the number returns to zero. |
The two angles
- Alpha angle between the upstroke and the plateau: about 100–110 degrees; it widens with expiratory airway obstruction.
- Beta angle between the plateau and the downstroke: about 90 degrees; it widens with rebreathing.
The number alone is not enough: the numeric value is a single point at end-expiration; the waveform is the whole story breath by breath. Always read the number together with the waveform.
Episode 3 — The screen says forty… the blood says fifty-five
The exhaled number is not the blood number
The normal EtCO₂ range is 35–45 mmHg, but that is end-tidal carbon dioxide, not arterial (PaCO₂).
Why is the blood higher?
Not every alveolus that receives gas receives enough blood, so its gas leaves without carbon dioxide and dilutes the number. This is alveolar dead space. In a healthy person PaCO₂ exceeds EtCO₂ by only about 2–5 mmHg.
When does the gap widen?
When dead space increases: pulmonary embolism, low cardiac output and hypotension, or chronic lung disease (COPD). In these cases you may see forty on the screen and feel reassured while the blood is far higher.
Clinical application
- Stable patient: capnography gives a close picture of the blood.
- Unstable patient or lung disease: match the number against an arterial blood gas (ABG) once, then follow the trend on the waveform.
- The gap is not fixed: if the patient bleeds or blood pressure falls, the screen number drops because blood is not reaching the lungs. Here capnography tells you about the circulation, not just breathing.
Episode 4 — Slow and tall… fast and low
| Comparison | Hypoventilation | Hyperventilation |
|---|---|---|
| Wave shape | Few, long and tall | Many, short and low |
| Number | Rises above 45 mmHg | Falls below 35 mmHg |
| Common causes | Respiratory depressants (opioids and sedatives), residual neuromuscular block, deep anesthesia, or low ventilator settings. Sometimes increased carbon dioxide production, as in fever or malignant hyperthermia. | Pain, anxiety, light anesthesia, or a ventilator set to a larger volume or rate than needed. It may be intentional as a temporary measure in some cases of raised intracranial pressure. |
The shape is the same square in both; what changed is height and rate.
Watch out
- Shallow breathing can give a low number despite high blood carbon dioxide, because a small breath is mostly dead space (hypopneic hypoventilation).
- A low number is not always hyperventilation: if the number falls and breathing has not sped up, think circulation (hypotension, low cardiac output, pulmonary embolism) and technical causes (leak around the tube, or a leaking sampling line).
- The right question is always: the number changed… but did the shape change? Did the respiratory rate change?
Coming episodes
Season two (the alarms): the shark fin, the baseline that does not return to zero, the line that became zero, and trachea or oesophagus. Added here as they are published.
Series glossary
| Term | Meaning |
|---|---|
| Capnography | Continuous measurement of exhaled carbon dioxide |
| Capnogram | The waveform drawn for each breath |
| EtCO₂ – End-tidal CO₂ | Carbon dioxide at the end of expiration |
| PaCO₂ | Partial pressure of carbon dioxide in arterial blood |
| SpO₂ | Peripheral oxygen saturation |
| FRC | Functional residual capacity, the lung's main oxygen store |
| Dead space | Gas that takes no part in gas exchange |
| Oxygenation / Ventilation | Oxygenation: oxygen reaching the blood · Ventilation: air moving in and out of the lungs |
| Hypercapnia / Hypocapnia | High / low carbon dioxide in the blood |
References
- Dr. Ahmed Shihab. «الموجة تحچي» (Capnography Decoded) – MAQ Academy educational series, episodes 1–4, October–November 2026. Instagram @dr_ahmed_shihab83.
- American Society of Anesthesiologists. Standards for Basic Anesthetic Monitoring (exhaled carbon dioxide monitoring during moderate and deep sedation).
Educational article for healthcare professionals. Clinical decisions rest with the anesthesiologist responsible for the patient and their hospital protocols.
Frequently asked questions
Why does SpO₂ stay normal despite apnoea?
Because a patient receiving supplemental oxygen has an oxygen store in the functional residual capacity, so desaturation is delayed. Capnography reads breath by breath and detects apnoea before saturation falls.
What is the normal EtCO₂ range?
35–45 mmHg, about 2–5 mmHg below arterial PaCO₂ in a healthy person. The gap widens with pulmonary embolism, low cardiac output and chronic lung disease.
What does a falling EtCO₂ with an unchanged respiratory rate mean?
Not necessarily hyperventilation. Think circulation – hypotension, low cardiac output or embolism – and technical causes such as a leak around the tube or in the sampling line.